Setup range for watt and Rflux measurements:
Specify the start and end wavelengths for the range
computation of the total power. The default range
is 400-700nm.
//////////////////////////////////////////////////////////////////
View -> Ref spectra:
Displays the reference and dark data previously saved as
ref(n)-dark(n). This is useful for troubleshooting absorbance
or transmission applications. The Ref spectra must not be
clipped at its peaks, indicating a bad reference.
//////////////////////////////////////////////////////////////////
View -> Channel:
Allows display selection for logical units when more than
1 spectrometer is installed. If you have multiple spectro-
meters this selection allows you to view data from the diff-
erent optical channels. Note: use "Setup -> Spectrometer
channels" to map different physical channels into the logical
unit selections (1-8).
//////////////////////////////////////////////////////////////////
View -> Multi-Graphs:
Allows display of multiple channels in a single graph.
Re-select to turn this feature off. The x-axis labeling can
be switched by using view -> channel -> n. All active
channels will be displayed.
//////////////////////////////////////////////////////////////////
View -> SnapShot:
Allows user to freeze display. Re-selecting SnapShot when the
display is frozen will re-start normal realtime display.
//////////////////////////////////////////////////////////////////
View -> Wave Ratio:
Allows 2 wavelengths to be selected to be displayed as a
ratio in the upper left hand corner. The default values
260/280 are used for DNA concentration measurement.
//////////////////////////////////////////////////////////////////
View -> Y scale:
Set Max Y:
Allows the user to specify the Y axis scale top value.
Selecting Y-axis rescale icon (left click) will override
the Max Y via autoscale. Entering a value of 0 will also
perform an autoscale.
as Log:
Converts the Y-axis into a Log base10 scale. Re-selecting
will convert back to a linear Y-axis.